5-29

Servicing

Replacing Instrument Assemblies

6

Be sure to cover the exposed end of the fiber-optic cable.

7

Use a T-10 TORX driver to remove the four screws securing the A2 assembly
to the instrument’s right-side frame.

8

Place the instrument on it’s right side.

9

Use a T-10 TORX driver to remove the three screws securing the A2 assembly
to the bottom of the instrument’s main deck sheet metal assembly.

10

Remove the A2 assembly.