Step 2. check the fuse, Step 2. check the fuse 1-5 – Agilent Technologies 83438A User Manual

Page 15

Advertising
background image

1-5

Getting Started

Step 2. Check the Fuse

Step 2. Check the Fuse

1

Locate the line-input connector on the instrument’s rear panel.

2

Disconnect the line-power cable if it is connected.

3

Use a small flat-blade screwdriver to open the pull-out fuse drawer.

4

Verify that the value of the line-voltage fuse in the pull-out drawer is correct.
The recommended fuse is an IEC 127 5

×20 mm, 6.3A, 250 V, Agilent part

number 2110-0703.

Notice that an extra fuse is provided in a drawer located on the fuse holder.

W A R N I N G

For continued protection against fire hazard, replace line fuse only
with same type and ratings, (type T 6.3A/250V for 100/240V
operation). The use of other fuses or materials is prohibited.