Setmea – Vaisala WAC155 User Manual
Page 42
User's Guide ______________________________________________________________________
40 ___________________________________________________________________M210822EN-A
Example:
>setmes 1 type mwvquery
0 TYPE=NONE INTERVAL=0.00
1 TYPE=MWVQUERY INTERVAL=1.00
2 TYPE=NONE INTERVAL=0.00
3 TYPE=NONE INTERVAL=0.00
>
SETMEA
The SETMEA command configures measurement parameters. The
configurable parameters include anemometer transfer function
parameters gain and offset.
SETMEA [allowedspdchange] [maxinvalidspdcount] [gaincorr]
[offset] [powersave] <value><cr>
where
allowedspdchange = Maximum allowed difference between two
adjacent samples in the sensor data, range
0.1 ... 25.0 (default 10.0)
maxinvalidspdcount = Maximum number of adjacent discarded
samples in the sensor data. Sensor data sanity
check functionality is enabled if this value is
not zero. Range 0 ... 15 (default 2)
gaincorr
= Anemometer transfer function gain parameter.
Range 0.0 … 10.0 (default 0.09853)
offset
= Anemometer transfer function offset
parameter. Range -10.0 … 10.0 (default
0.4054)
powersave
= Defines whether sensor power is pulsed or
provided continuously. Pulsing saves power,
since the sensors are only powered when they
are being read.
0 = continuous power
1 = pulsed power (default)
value
= Defines the value for the option.