Appendix h. calculating multipliers and offsets, Appendix h – Campbell Scientific AL200 ALERT2 Encoder, Modulator, and Sensor Interface User Manual

Page 41

Advertising
background image

Appendix H. Calculating Multipliers and

Offsets

Unlike ALERT which could only represent measurement values between 0 and

2047, ALERT2 has the ability to represent a much larger range of floating

point numbers. It is no longer necessary to scale a reading to a unit-less

number less than 2047. Instead, with ALERT2, you can transmit values in

engineering units with a high degree of resolution. The simplified example

below demonstrates how to calculate the multiplier and offset that are needed

to convert a pressure transducer’s output, 0-5 volts, to feet of water.

For this example we will use a pressure transducer that has a 0 to 30 psi range

and a 0 to 5 volts output. From the manufacturer supplied calibration report it

is known that the sensor outputs 0.0123 volts at 0 psi and 4.987 volts at 30 psi

(12.3 and 4987 millivolts respectively).

The relationship between the sensor’s output in millivolts and pressure is

defined by the equation of a line (assuming linearity), Y = mX + b, where Y is

pressure, m is the calculated multiplier, X is the sensor’s output in millivolts,

and b is the calculated offset.

Y = mX + b

OR

Pressure = Multiplier • Sensor Output + Offset

The multiplier is determined by dividing the sensor’s measurement ranges by

the sensor’s range of output. For this example the multiplier will be:

1

X

2

X

1

Y

2

Y

Multiplier

=

OR

mV

per

psi

0060

.

0

)

7

.

4974

(

30

)

3

.

12

4987

(

0

30

Multiplier

=

=

=

The offset is determined by entering the multiplier into the pressure equation

and solving for the offset using a known pressure and sensor output.

Offset = Pressure – (Multiplier • Sensor Output)

OR

(

)

mV

per

psi

074

.

0

3

.

12

0060

.

0

0

Offset

=

=

H-1

Advertising