Athena Technologies 6050 User Manual

Page 10

Advertising
background image

THERMOCOUPLE PLACEMENT

Proper thermocouple placement can eliminate many problems in

the system. The probe should be placed so that it can detect

any temperature change with minimal thermal lag. In a process

that requires fairly constant heat output, the probe should

be placed close to the heater. In processes where the heat

demand is variable, the probe should be close to the work

area. Some experimenting with probe location is often needed

to find its optimum position.

In

a bath process, addition of a stirrer will help to

eliminate thermal lags. Since the thermocouple is basically

a point measuring device, putting more than one

in parallel will provide an average temperature

produce better results in air heated processes.

thermocouple

reading and

NOTE: If controls with "F" or "S" outputs drive

grounded or hot input terminals (not floating),

thermocouple must be used. Otherwise, when both

loads with

an isolated

input and

output are grounded, ground loop currents will result,

causing errors and controller damage.

Standard thermocouple limits of error are 4°F or 0.75% of

sensed temperature (half that for special) plus drift caused

by improper protection or overtemperature. This is far

greater than controller error, and can not be corrected at

the sensor except by selection and replacement.

Start-Up

Before line voltage is applied, double-check the following:

Make sure the thermocouple type is correct, and

properly connected (see section on thermocouples)

to terminals 1 and 2 (red on 2).

Make sure there is no AC or DC voltage leading or

arcing to T/C.

Make sure the proper terminals are selected for

line voltage. (8 & 9 for 12OV, 8 & 10 for 24OV)

Check to assure there are no heater shorts, or

shorts to ground, and no bare wires or frayed insulation.

Make sure correct plug-in module is used.

9

Advertising