Section 2 Installation

2.3.3 Important Note

2.4.1 CM2 Sensors

Nominal

CM2

Width

Actual

CM2

Width

L

S

R

H

M

Weight

Each

in. mm. in. mm. in.

mm. in. mm. in. mm. in. mm. in.

mm. lb kg.

8.0 200 10.0 254 40.0 1016 8.0 203 14.5 370 2.50 64

6.0 152 55 25

24.0 600 26.0 660 60.0 1520 25.0 610 18.0 460 3.12 80

12.0 305 186 85

40.0 1000 42.0 1066 80.0 2030 40.0 1020 20.0 510 3.50 89

24.0 610 355 165

When using an ultrasonic level transducer, or a hydrostatic level transducer in a stilling well, it
must be installed at least 1 channel width before or after the ends of the ramps when the
ChannelMag sensor lies on the channel bed. This avoids surface level drops directly above the
sensors caused by sub-critical flows, or level rises caused by super critical flows.

These critical flows are circumvented if the CM2 is let into a recess in the bed. As such, the upper
surface of the CM2 sensor may be up to 1” (25mm) above the channel bed to avoid silt build-up.
The ultrasonic level transducer may then be mounted directly above the CM2 sensor(s). Note the
minimum dead band for the ultrasonic level transducer.

When a hydrostatic level transducer is used it is normally installed in the ramp, with its sensing at

the point of the ramp. This also avoids critical flow drops or rises in the media level above the

ChannelMag sensor.

2.4 Installation,

Dimensions and

Approx. Weights

Pipe Confi

guration

Single Sensor

Number of Straight

Pipe Diameters D

Upstream

Downstream

Multiple Sensors

Number of

Straight Pipe Diameters D

Upstream Downstream

90° bend

upstream and

downstream

10 D

5 D

5 D

2 D

After a tee

15 D

5 D

8 D

2 D

Upstream

partially

closed valve

25 D

5 D

15 D

2 D

Downstream

partially

closed valve

8 D

5 D

Note