Myron L TH1 User Manual

Page 14

Advertising
background image

10

3.

Use the

or

key to select type of solution desired

(ref. Solution Characteristics, pg. 28). The selected solution

type will be displayed: KCl, NaCl or 442.

4.

Press

to accept new solution type.

VII.

CALIBRATION

A. Calibration Intervals

Generally, calibration is recommended about once per month with

Conductivity or TDS solutions. Calibration with pH solutions should be

checked twice a month. (ref. CALIBRATION INTERVALS, pg. 14).

B. Rules for Calibration of the TechPro II

1. Calibration Steps

a. Starting Calibration

Calibration is begun by pressing

while measuring Conductivity,

TDS or pH. Measuring continues, but the CAL icon is on, indicating

calibration is now changeable.

The reading is changed with the

and

keys to match the

known value. The calibration for each of the 3 solution types may be

performed in either conductivity or TDS mode.

Depending on what is being calibrated, there may be 1, 2 or 3 steps to

the calibration procedures.

Once in “CAL” mode, the

key becomes an “ACCEPT” key. At each

point, pressing

accepts the new calibration value and steps you to

the next adjustment (or out of CAL mode if there are no more steps).

In the first six sections, you have learned all

you need to take accurate measurements.

The following sections contain calibration,

advanced operations and technical information.