26

C. An Example of 2 different solution selections and the

resulting compensation

How much error results from treating natural water as if it were KCl at

15°C?

A tap water solution should be compensated as 442 with a tempco of

1.68 %/°C, where the KCl value used would be 1.90 %/°C.

Suppose a measurement at 15°C/59°F is 900 microsiemens of true

uncompensated conductivity.

Using a 442 correction of 10 (degrees below 25) x 1.68% indicates the

solution is reading 16.8% low. For correction, dividing by (.832) yields

1082 microsiemens as a compensated reading.

A KCl correction of 10 (degrees below 25) x 1.9% indicates the solution

is reading 19% low. Dividing by (.81) yields 1111 microsiemens for a

compensated reading. The difference is 29 out of 1082, or 2.7%.

D. A Chart of Comparative Error

In the range of 1000 µS, the error using KCl on a solution that should be

compensated as NaCl or as 442, is illustrated in Chart 2 on pg. 27.