Myron L TH1 User Manual

Page 27

Advertising
background image

23

Corrective Action

Check connections or replace battery

(ref. Battery Replacement, pg. 20).

1. Recalibrate instrument.

2. Thoroughly rinse sensor well.

3. Recalibrate using fresh buffers (ref. pH

Buffer Solutions, pg. 24).

Replace pH sensor (ref. pH Sensor

Replacement, pg. 20).

Clean and rejuvenate sensor (ref. Cleaning

Sensors, pg. 20) and recalibrate. If no

improvement, replace pH sensor

(ref. pH Sensor Replacement, pg. 20).

Clean and rejuvenate sensor (ref. Cleaning

Sensors, pg. 20) and recalibrate. If no

improvement, replace pH sensor

(ref. pH Sensor Replacement, pg. 20).

Clean cell cup and electrodes

(ref. Cleaning Sensors, pg. 20).

Clean cell cup and electrodes

(ref. Cleaning Sensors, pg. 20).