Myron L TH1 User Manual

Page 17

Advertising
background image

13

will flash, and the TechPro II will not adjust. The uncalibrated pH value

displayed in step 4 will assist in determining the accuracy of the pH sensor.

If the pH reading is above 8 with pH 7 buffer solution, the sensor well needs

additional rinsing or the pH sensor is defective and needs to be replaced.

5.

Press

or

until the display reads 7.00.

NOTE: Attempted calibration of >1 pH point from factory calibration will

cause “FAC” to appear. This indicates the need for sensor replacement

(ref. Troubleshooting pg. 22) or fresh buffer solution. The “FAC” internal

electronic calibration is not intended to replace calibration with pH buffers.

It assumes an ideal pH sensor. Each “FAC” indicates a factory setting for

that calibration step (i.e., 7, acid, base).

You may press

to accept the preset factory value, or you may

reduce your variation from factory setting by pressing

or .

6.

Press

to accept the new value. The pH Zero Calibration

is now complete. You may continue with pH Gain Calibration or

exit by pressing any measurement key.

b. pH Gain Calibration (TPH1 & TH1)

Important: Always calibrate or verify your TechPro II with a pH 7 buffer

solution before adjusting the gain with acid or base buffers, i.e., 4 and/or

10, etc. Either acid or base solution can be used for the 2nd point “Gain”

calibration and then the opposite for the 3rd point. The display will verify

that a buffer is in the sensor well by displaying either “Acd” or “bAS”.

1.

The pH calibration mode is initiated by either completion of the

pH Zero Calibration, or verifying 7 buffer and pressing the

key twice while in pH measurement mode.

2.

At this point the “CAL”, “BUFFER” and “Acd” or “bAS

annunciators will be displayed (see Figures 4 and 5).

Figure 4

BUFFER

pH

CAL

Figure 5

pH

BUFFER

CAL

Advertising
This manual is related to the following products: