WIKA DPT-10 User Manual

Page 64

Advertising
background image

+

A

B

6

7

3

2

4

1

5

+

II

I

III

III

Fig. 47: Preferred installation for closed vessels

I

DPT10

II

3

-fold valve block

III

Precipitator

1

, 5 Drain valves

2

, 4 Inlet valves

6

, 7 Vent valves on DPT10

A, B Blocking valves

Proceed as follows:

1

Fill the vessel to just above the lower tap

2

Fill measuring system with medium

Close valve 3: Separate plus and minus side

Open valve A and B: Open block valves

3

Vent plus side (probably empty minus side)

Open valve 2 and 4: Discharge medium on the plus side

Briefly open valve 6 and 7, then close again: Fill the plus side
completely with the medium and remove air.

4

Set measurement loop to operation

Now:

Valve 3, 6 and 7 are closed

Valves 2, 4, A and B are open

Then carry out adjustment, see below.

Installation for closed
vessels

Prepare the adjustment

64

Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA

8 Set up

37245

-EN

-100120

Advertising