WIKA DPT-10 User Manual

Page 74

Advertising
background image

Max. torque screws mounting strap

30 Nm

Max. torque screws socket external housing

5 Nm (3.688 lbf ft)

Weight approx.

4.2 … 4.5 kg (9.26 … 9.92 lbs), depending on
process fitting

Output variable

Output signal

digital output signal, format according to IEEE-754

Sensor address

126 (default setting)

Current value

10 mA, ±0.5 mA

Dynamic behaviour output

Run-up time

20 s

I

63 %

100 %

t

t

1

t

2

Fig. 51: Presentation of the dead time t

1

and the time constant t

2

Version, nominal measuring range

Dead time t

1

Time constant t

2

Basic version 10 bar and 30 bar

100 ms

250 ms

Basic version 100 mbar

100 ms

180 ms

Basic version 500 mbar

100 ms

180 ms

Basic version, 3 bar

100 ms

180 ms

Basic version 16 bar and 40 bar

100 ms

180 ms

Isolating diaphragm version, all nominal measuring
ranges

Depending on the isolat-
ing diaphragm

Depending on the isolating
diaphragm

Response time with Profibus PA

-

cyclically approx.

10 ms

-

acyclically approx.

50 ms

Damping (63 % of the input variable)

0 … 999 s, adjustable

74

Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA

11 Supplement

37245

-EN

-100120

Advertising