6 ttl i/o directional control input configuration, 7 pump motor operating ttl output configuration – Next Advance SP2200 User Manual

Page 15

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe

Pump

Publication #1200-02

10

04/29/10

6.5 TTL I/O Operational Trigger Default Configuration

Command: "TRG <aa>", <aa> as below in Settings.

Configures the default mode of the TTL I/O ‘Operational Trigger’ (pin 2) that will control the Pumping
Program’s operation. This default setting can be overridden within a Pumping Program. (See sec. 9.1, TTL I/O
Operational Controls
). The 2 letter configuration Setting parameter to the right of the colon (:) is defined in
column ‘Setting’ as follows:

Setting

<aa>

RS-232 Command Setting

FUN TRG <n>

<n> =>

Name

Function

FT

0

Foot Switch

Falling edge starts or stops the Pumping Program

FH

1

Foot Switch Hold

Falling edge starts the Pumping Program
Rising edge stops the Pumping Program

F2

2

Foot Switch
Reversed

Rising edge starts or stops the Pumping Program

LE

3

Level Control

Falling edge stops the Pumping Program
Rising edge starts the Pumping Program

ST

4

Start Only

Falling edge starts the Pumping Program

T2

5

Start Only Reversed

Rising edge starts the Pumping Program

SP

6

Stop Only

Falling edge stops the Pumping Program

P2

7

Stop Only Reversed

Rising edge stops the Pumping Program

6.6 TTL I/O Directional Control Input Configuration

Command: "DIN <n>"

Configures how the TTL input ‘Pumping Direction’ (pin 3) will control the pumping direction. (See sec. 9.1,
TTL I/O Operational Controls).

Setting <n>

Mode

Function

0

Reciprocating Pumps

Falling edge: Infuse; Rising edge: Withdraw
- Used for 2 pumps operating in opposite directions

1

Dual Pump

Falling edge: Withdraw; Rising edge: Infuse
- Used with 2 pumps operating in the same direction

6.7 Pump Motor Operating TTL Output Configuration

Command: "ROM <n>"

Configures the functionality of the 'Pump Motor Operating' TTL output pin (TTL pin 7).

Setting <n> Mode

0

Sets the output to logic high only when the motor is operating (pumping).

Sets the output to logic low when the motor is not operating or when the Pumping Program is executing
a pause timer or is stopped

1

Sets the output to logic high when the motor is operating (pumping) or when the Pumping Program is
executing a pause timer.

Sets the output to logic low when the Pumping Program is stopped