Next Advance SP2200 User Manual

Page 24

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe

Pump

Publication #1200-02

19

04/29/10

Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked back
volume of the previous dispense. By changing the last Phase to a [FUN LOP nn] function, the total number of
dispenses can be set.

When entering a function with associated data, such as with the ‘Pause’ in Phase 5 or the ‘Loop’ in Phase 6, the
function is entered in 2 steps. First select the function and store it. Then enter the associated data.

Phase

Function

Rate

Volume

Direction

1

RAT

750 ml/hr

2.0 ml

Infuse

Phase

Function

Rate

Volume

Direction

2

RAT

750 ml/hr

0.25 ml

Withdraw

Phase

Function

3

LPS

Phase

Function

4

LPS

Phase

Function

5

PAS 90

Phase

Function

6

LOP 03

Phase

Function

7

BEP

Phase

Function

8

PS:30

Phase

Function

Rate

Volume

Direction

9

RAT

750 ml/hr

2.25 ml

Infuse

Phase

Function

Rate

Volume

Direction

10

RAT

750 ml/hr

0.25 ml

Withdraw

Phase

Function

11

LPE

7.4.3 Example 3: Ramping the Flow Rate

m

l/h

r

R

at

e

Time

0

150

200

250

Phases
2 - 4

Phases
5 - 8

Phases
9-11

Note: Graph is representative only.
Pumping rate increments and decrements in steps.

Continuously ramp up and down the pumping rate. Starting at 200 ml/hr, the pumping rate will increment to
250 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Then the pumping rate will decrement to
150 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Finally, the pumping rate is incremented
back to 200 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed, then the process is repeated.