4 pumping program examples – Next Advance SP2200 User Manual

Page 23

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe

Pump

Publication #1200-02

18

04/29/10

7.4 Pumping Program Examples

7.4.1 Example 1: 2 Step Rate

ml Dispensed

0 5

30

0

2.5

500

m

l/

h

r

R

a

te

Infuse 5.0 ml at 500 ml/hr, and then infuse 25.0 ml at 2.5 ml/hr. Then stop the pump.

Phase

Function

Rate

Volume

Direction

1

RAT

500 ml/hr

5.0 ml

Infuse

Phase

Function

Rate

Volume

Direction

2

RAT

2.5 ml/hr

25.0 ml

Infuse

Phase

Function

3

STP

To send this program to the pump, send the following sequence of commands:

PHN 1
FUN RAT
RAT 500 MH
VOL 5.0
DIR INF

PHN 2
FUN RAT
RAT 2.5 MH
VOL 25.0
DIR INF

PHN 3
FUN STP

7.4.2 Example 2: Repeated Dispenses with Suck Back

m

l/

h

r

R

at

e

Time

Advertising