Appendix, 1 rs-232 command summary – Next Advance SP2200 User Manual

Page 46

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe

Pump

Publication #1200-02

41

04/29/10

10. Appendix

10.1 RS-232 Command Summary

<command> =>
DIA [ < float > ]

Syringe inside diameter

PHN [ < phase data > ]

Program Phase number

FUN [ < phase function > ]

Program Phase function

< phase function > =>

RAT

Pumping rate.

INC

Increment rate.

DEC

Decrement rate.

STP

Stop pump.

PAS <number data>

Pauses pumping for ‘nn’ seconds, if 0, wait for start trigger.

PAS [n.n]

Pauses pumping for 'n.n' seconds.

JMP <phase data>

Jump to Program Phase.

PRI

Sub-Program Selection Input from expansion connector

PRL <number data>

Sub-Program Start Label definition.

LOP <count data>

Loop to previous loop start ‘nn’ times.

LPS

Loop starting Phase.

LPE

Loop end Phase.

IF <phase data>

If Program input low, jump to Program Phase.

EVN <phase data>

Set event trigger.

EVS <phase data>

Set event square wave trigger.

EPL <expansion pin>

Pause program, wait for low edge or level on expansion pin

EPE <expansion pin>

Pause program, wait for falling edge on expansion pin

EVE <expansion pin>

Re-direct event trigger to expansion pin

EVR

Event trigger reset.

OUT { 0 | 1 }

Set programmable output pin.

OE0 <expansion pin>

Set expansion pin low

OE1 <expansion pin>

Set expansion pin high

TRG <n>

Override Operational Trigger default setting

BEP

Sound short beep.

RAT {C | I ] [ <float> [ UM | MM | UH | MH ]] Pumping rate
VOL [ <float> | <volume units>]

Volume to be Dispensed, or set Volume units

DIR [ INF | WDR | REV ]

Pumping direction

RUN [ <phase data> ]

Starts the Pumping Program

[E [<phase data>] ]

Pumping Program event trigger

STP

Stop/pauses the Pumping Program

DIS

Query volume dispensed

CLD { INF | WDR }

Clear volume dispensed

SAF [ <n> [ <n> [ <n> ] ] ]

Safe communications mode

AL [ <on-off> ]

Alarm mode

PF [ <on-off> ]

Power failure mode

TRG [ FT | FH | F2 | LE | ST | T2 | SP | P2]

Operational trigger default mode

DIN [ 0 | 1]

Directional input control mode

ROM [0 | 1]

Pump Motor Operating TTL output mode

LOC [ <on-off> ]

Lockout mode for Sub-Program Auto-Run

OUT [ E <expansion pin> | 5] { 0 | 1 }

Set TTL output level

IN { E <expansion pin> | 2 | 3 | 4 | 6 }

Query TTL input level

BUZ [ 0 | { 1 [ < n > ] } ]

Buzzer control

VER

Query firmware version

*ADR [ <n> [<n>] [ B <baud-rate>] ]

Network address and baud rate (system command, valid
regardless of current address)