Configuring for remote control, Gpib interface configuration, Lan interface configuration – Agilent Technologies E8257D/67D User Manual

Page 14

Advertising
background image

10

Chapter 2

Getting Started
Configuring for Remote Control

Configuring for Remote Control

GPIB Interface Configuration

1. Press

Utility

>

GPIB/RS-232 LAN

>

GPIB Address

.

2. Use the numeric keypad to set the desired address and press

Enter

.

The arrow keys or the front panel knob can be used to set the desired address.

The signal generator’s GPIB address is set to 19 at the factory. The acceptable range of addresses is 0
through 30. The GPIB address is a persistent state; it is not affected by a signal generator preset or by a
power cycle.

LAN Interface Configuration

Use one of the following procedures to assign a hostname and IP address to the signal generator. The
hostname and IP address are persistent states; they are not affected by an instrument preset or a power cycle.

NOTE

Verify that the signal generator is connected to the LAN using a 10BASE-T LAN cable.

Manual Configuration

1. Press Utility > GPIB/RS-232 LAN > LAN Setup

.

2. Press

Hostname

.

NOTE

Hostname

field will only be available when the

DHCP

is off (

Manual

mode).

3. Use the labeled text softkeys and/or numeric keypad to enter the desired hostname.

To erase the current hostname, press

Editing Keys

>

Clear Text

.

4. Press

Enter

.

5. Set

LAN Config Manual DHCP

to Manual.