Table 3-6 – Agilent Technologies E8257D/67D User Manual

Page 34

Advertising
background image

30

Chapter 3

Operation Verification
Checking the Output Power

Table 3-5

Leveled Output Power Limits, E8257D cont.

E8257D

Option 550

Option 567

Frequency

Amplitude

Limits

Frequency

Amplitude

Limits

Option

dBm

dB

Option

dBm

dB

250 kHz

None

9

±0.6

250 kHz

None

9

±0.6

1EA

15

±0.6

1EA

15

±0.6

1E1

7

±0.6

1E1

7

±0.6

1E1 & 1EA

14

±0.8

1E1 & 1EA

14

±0.8

3.2 GHz

None

9

±0.8

3.2 GHz

None

9

±0.8

1EA

15

±0.8

1EA

15

±0.8

1E1

7

±0.8

1E1

7

±0.8

1E1 & 1EA

14

±0.8

1E1 & 1EA

14

±0.8

20 GHz

None

9

±0.9

20 GHz

None

9

±0.9

1EA

18

±1.0

1EA

18

±1.0

1E1

7

±0.9

1E1

7

±0.9

1E1 & 1EA

16

±1.0

1E1 & 1EA

16

±1.0

40 GHz

None

9

±0.9

40 GHz

None

9

±0.9

1EA

14

±1.0

1EA

14

±1.0

1E1

7

±0.9

1E1

7

±0.9

1E1 & 1EA

12

±1.0

1E1 & 1EA

12

±1.0

50 GHz

None

5

±1.2

50 GHz

None

5

±1.2

1EA

14

±1.2

1EA

14

±1.2

1E1

3

±1.2

1E1

3

±1.2

1E1 & 1EA

10

±1.2

1E1 & 1EA

10

±1.2

Advertising