Agilent Technologies E8257D/67D User Manual

Page 32

Advertising
background image

28

Chapter 3

Operation Verification
Checking the Output Power

Perform the following procedure to check the signal generator output power:

1. Zero and calibrate the power sensor to the power meter as shown:

2. Connect the equipment as shown:

3. Press

Preset

.

4. Press

RF On/Off

. The

RF On

annunciator is now displayed.

5. Press

Mod On/Off

. The

Mod Off

annunciator is now displayed.

6. Press

Frequency

and enter the first frequency value listed in

Table 3-1

.

7. Press

Amplitude

and enter the amplitude value for your option and frequency value.

8. Measure the power level.

9. Repeat step

6

through

8

and measure the power levels for each remaining frequency listed in

Table 3-4

thru

Table 3-6

. Record any power levels not within the acceptable limits.

Advertising