24

Chapter 3

Operation Verification
Checking the Maximum Leveled Power

NOTE

For Option 1EH and Option UNW, use the data sheet to determine low band power level
limits.

E8257D

Option 520

With Option 1E1

11 dBm

15 dBm

250 kHz

3.2 GHz

50

11 dBm

18 dBm

3.21 kHz

20 GHz

200

E8257D

Option 540

With Option 1E1

7 dBm

14 dBm

250 kHz

3.2 GHz

50

7 dBm

16 dBm

3.21 kHz

20 GHz

200

7 dBm

12 dBm

20.1 GHz

40 GHz

200

E8257D

Option 550 and

Option 567

With Option 1E1

3 dBm

13 dBm

250 kHz

3.2 GHz

50

3 dBm

14dBm

3.21 kHz

20 GHz

200

3 dBm

10 dBm

20.1 GHz

60 GHz

200

3 dBm

8 dBm

60.1 GHz

67 GHz

100

Table 3-1

Frequency and Power Level Limits