Power plugs and cord, Power plugs and cord 2 – Agilent Technologies 5975 User Manual

Page 26

Advertising
background image

26

Site Preparation Guide

Site Preparation

Power plugs and cord

The MSD is supplied with a power cord and plug appropriate for the country
from which the order originates. For example, if an order originates from a
Agilent Technologies sales office in Germany, the power cord and plug
supplied will be compatible with the standard voltage and outlet configuration
in Germany. See “Power Cords" on page 41 for illustrations of the power cords
available.

Data system components also include power cords with plugs appropriate for
the country where the order was placed.

Power cord lengths for the MSD and the data system components and
accessories are approximately 2.3 m (7.5 ft).

ChemStation computer

100-120 or 220-240 VAC+5/-10%,

50/60 Hz ± 5%

1000 VA

15 A

3 to 5

* The MSD operates on either voltage range. The foreline and turbo pumps are supplied for the voltage range suitable to the

country where the order originates

† Americas 120 V requires 20 A dedicated line. Americas 240 V requires 15 A dedicated line.

‡ Option 003, fast oven, is a unit with operating range 198 to 231 V. Most laboratories have 4-wire service resulting in 208 V at

the wall receptacle. It is important to measure the line voltage at the receptacle for the GC.

** Requires an isolated ground and dedicated outlet.

Table 4

Power requirements (continued)

Product

Line voltage

Maximum continuous
power consumption

Supply
circuit rating

Outlets
required

C A U T I O N

If an instrument is being ordered from one location but is to be installed in another
location with different electrical power characteristics, this must be noted on the order.
A special note must also be made if the electrical power at the site is different from the
standard electrical power in that country.

WA R N I N G

Make sure the power cords supplied with the MSD are appropriate for your country
and site before installing the instrument. See “Power Cords” on page 41.

5975 Series Site Prep.book Page 26 Thursday, January 31, 2008 9:47 AM