Site Preparation Guide

Site Preparation

Air Conditioning and Environmental Requirements

Air conditioning considerations include temperature, humidity, airborne dust,
and exhaust venting. Each of these is considered in more detail in the
following material.

Temperature, humidity, and altitude

The 5975 Series MSD is specified for operation under the following conditions:

Operation requires constant temperature

(variations <2

°C/h)

Operation and storage require a noncondensing, noncorrosive atmosphere

The temperature and humidity limits in Table 5

Maximum altitude for operation: 2300 m (7500 ft)

Environmental control systems must maintain these temperature and
humidity ranges.

The MSD is rated for 1200 Watts (4000 BTU/h). The GC and the data system
also contribute significantly to the cooling load although the exact amounts
depend on their configurations. Additional allowances should be made for
other heat sources such as heat from other equipment, heat from adjacent
rooms, and heat from laboratory personnel.

Table 5

Temperature and humidity limits

Temperature

Humidity (relative)

Operation

15 °C to 35 °C (59 °F to 95 °F)

40% to 80%

Storage

–20 °C to 70 °C (–4 °F to 158 °F)

0% to 95%