Site Preparation

Site Preparation Guide

27

Other electrical considerations

Additional electrical considerations include:

Electromagnetic interference (EMI), such as is generated by NMRs, radio

transmitters, and microwave links, may interfere with system performance.

Protect the system from static electricity by observing humidity and

temperature requirements. Minimize the presence of nonconductive
products such as carpets and vinyl floor tiles.

Install emergency-off pushbuttons that can disconnect power to the

ventilation system and all electric equipment in the room except overhead
lighting.

Provide separate convenience outlets for building maintenance and other

appliances. Convenience outlets must be on circuits separate from the MSD
system. Convenience outlets must share the normal building distribution
ground, not the MSD system ground.

In some geographical areas it may be advisable to install lightning

protection for personnel and equipment.

WA R N I N G

Do not use extension cords with the MSD or GC. Extension cords normally are not
rated to carry enough power and can be a safety hazard.

N O T E

Maintain easy access to the power cords so they can be disconnected for maintenance.