Carrier and reagent gas requirements, Carrier and reagent gas requirements 3, Carrier and reagent gas requirements 31 – Agilent Technologies 5975 User Manual

Page 31

Advertising
background image

Site Preparation

Site Preparation Guide

31

Carrier and Reagent Gas Requirements

Table 6 lists the specifications for the carrier and reagent gases. Pressures and
flows are as measured at the bulkhead fittings on the back of the instrument,
not at the gas supply. We recommend that additional shut-off valves be
installed near the instruments.

Table 6

Carrier and reagent gas requirements

Purity

Typical pressure range,
kPa (psi)

Typical flow, mL/min

Helium (required)

99.9995% or better and
hydrocarbon-free

345 to 550 (50 to 80)

20 to 50 (column and split flow)

Methane reagent gas
(required for CI operation)

Research or SFC grade,

99.999%

103 to 172 (15 to 25)

1 to 2

Isobutane reagent gas
(optional)

Instrument grade,

99.99%

103 to 172 (15 to 25)

1 to 2

Ammonia reagent gas
(optional)

Research or SFC grade,
99.9995%

34 to 55 (5 to 8)

1 to 2

Carbon dioxide reagent gas
(optional)

SFC Grade,

99.995%

103 to 138 (15 to 20)

1 to 2

WA R N I N G

Methane is flammable and ammonia is corrosive and toxic. Read and follow the
safety information that comes with the gases.

5975 Series Site Prep.book Page 31 Thursday, January 31, 2008 9:47 AM