Appliance dimensions setting up, Do not cover ventilation openings or grille, Climate rating – Liebherr MKv 3910 User Manual

Page 4: Features

Advertising
background image

16

Appliance dimensions

Setting up

• Avoid positioning the appliance in direct sunlight or near cookers,

radiators and similar sources of heat.

• The floor on which the appliance stands should be horizontal

and level. Compensate for uneven floors with the adjustable feet.

Do not cover ventilation openings

or grille.

• Standard EN 378 specifies that the room in

which you install your appliance must have

a volume of 1 m

3

per 8 g of R 600a refriger-

ant used in the appliance, so as to avoid the

formation of inflammable gas/air mixtures in

the room where the appliance is located in

the event of a leak in the refrigerant circuit.

The quantity of refrigerant used in your ap-

pliance is indicated on the type plate on the

inside of the appliance.

Climate rating

The climate rating indicates the room

temperature at which the appliance may

be operated in order to achieve full refrig-

eration performance.
The climate rating is indicated on the

type plate.The position of the type plate

is shown in the section entitled Description of the appliance.

Climate rating

Room temperature

SN

+10°C to +32°C

N

+16°C to +32°C

ST

+16°C to +38°C

T

+16°C to +43°C

SN-ST

+10°C to +38°C

SN-T

+10°C to +43°C

Do not operate the appliance outside the

specified room temperature range.

Energy consumption

Temperature increase after a power failure

Model

MKv 3910

MKv 3913

Energy consumption to

DIN 58345

0.8 kWh/24h 1.5 kWh/24h

Temperature increase after

a power failure from +5°C to

+10°C

50 minutes

30 minutes

Features

- Audible and visual temperature alarm.
- Audible and visual door open alarm.
- Floating contact for connection to a remote monitoring system.
- Serial interface (RS485) for external temperature and alarm

documentation.

- Maximum/minimum interior temperatures are stored.
- Last 3 temperature alarms are saved with time, date and duration

of alarm.

- Last 3 power cuts are saved with time, date and duration of power

cut.

- Safety thermostat to avoid temperatures below +2°C.

It is essential to use these safety facilities to avoid damage

to stored items. These facilities must not be deactivated or

decommissioned!

Range of appliance use
The appliance is designed for storing and cool-

ing pharmaceuticals to DIN 58345 as long as the

following conditions are satisfied:
• The appliance only complies with DIN 58345 if

the battery of the power failure alarm integrated

in the appliance is fully charged (see section

entitled "Electrical connection").

• The appliances only satisfies the requirements

of DIN 58345 with its factory temperature set-

ting. If the temperature setting is adjusted (which

can only be done by the service technician), its

compliance with DIN 58345 will be void.

The appliance is not suitable for use in explosion-

hazard areas.