Alternative site testing – Bayer Breeze-2 User Manual

Page 23

Advertising
background image

9

www.BayerDiabetes.com

Unusual Test Results

If your blood glucose result doesn’t match the way you feel,

follow these steps:
1. run a control test (see page 20);

2. retest your blood glucose (see page 10);

3. call your doctor immediately if your result still does not agree

with the way you feel.

Alternative Site Testing

Blood samples for glucose testing may be taken from sites

other than your fingertips. Alternative site testing using blood

from the forearm or palm may give glucose results that signifi-

cantly differ from fingertip blood. Differences occur when blood

glucose levels are changing rapidly, such as after a meal, after

insulin and with exercise.
We recommend the following:
1. Consult your healthcare professional to determine if

alternative site testing is right for you.

2. Use a fingertip sample for testing blood glucose whenever

you will be participating in risky activities, such as before

driving a car or operating machinery.

3. Alternative site testing is not recommended if you have

hypoglycemic unawareness (you do not recognize the

symptoms of or cannot tell when you have low blood glucose).

Please consult with your healthcare professional if you have

low blood glucose levels.

4. Select a soft, fleshy area of skin that is free from hair and

visible veins for alternative site testing. Wash the site with

soap and warm water, then rinse and dry thoroughly.

5. Use alternative site testing for blood glucose tests only when

it is more than 2 hours after:

• a meal

• taking medication

• exercise

Advertising