Roubleshooting, Rror, Odes – Bayer Breeze-2 User Manual

Page 44

Advertising
background image

What You See

What It Means

What You Should Do

Test result is below

10 mg/dL.

Repeat the test and make sure

the test strip is completely filled.

If LO appears again, contact

your physician or healthcare

professional.

Test result is above

600 mg/dL.

Wash your hands and the

test site. Repeat the test. If

HI appears again, contact

your physician or healthcare

professional.

You have not pushed

the meter handle all

the way in.

Push the meter handle in until it

clicks. You must push the handle

all the way in and see the “apply

blood” prompts before you touch

the strip to the blood drop.

• The 10-test disc is

out of test strips.

• There is no disc in

the meter.

• The disc has been

inserted upside down.

• The disc is defective.

Insert a new 10-test disc. If a full

disc is already inserted, remove and

inspect for foreign material. Rotate

the disc and reinstall it, bumpy side

up. If error repeats, try inserting

a new disc. If the error continues,

contact Bayer Customer Service.

The meter will mark

a test result with a

thermometer and an

if the meter is too cold

or too hot when the test

is performed. The result

will not be included in

any averages calculated.

Test only when the meter

temperature is between 50°F

and 113°F (10°C and 45°C).

Bayer Customer Service: 1-800-348-8100

0

T

roubleshooTIng

and

e

rror

C

odes

:

When you have a problem or a question concerning what

you see on your screen, the following may be helpful.

Advertising