What You See

What It Means

What You Should Do

Control Test
result is out of
range (too high
or too low).

• Control solution is

past its expiration

date or is past the

6-month open

use-life date.

• 10-test disc is past its

expiration date.

• Deteriorated test

strip due to heat or

exposure to moisture.

• Control solution

is not at room

temperature.

• Not enough control

solution drawn into

test strip.

Run another control test. If the

result is still out of range, retest

with a new 10-test disc and

control solution with an acceptable

expiration date and open use-life

date. If still out of range, contact

Bayer Customer Service.

Meter begins
countdown before
blood is applied.

The inside of your

meter may be wet.

Wick the moisture away (see

page 34). If you continue to

experience difficulty, contact

Bayer Customer Service.

Some display
segments do not
appear to be
showing when

is held in or

the meter handle
is all the way out.

This may affect the way

you see your results.

If there is a difference in your

meter’s display and the one

shown on page 5, contact Bayer

Customer Service.

Bayer Customer Service: 1-800-348-8100