Bayer Breeze-2 User Manual

Page 47

Advertising
background image

What You See

What It Means

What You Should Do

or higher

Meter has detected an

error; accurate result is

not possible.

Turn the meter off, then on

using

, then retest. If the

error remains, contact Bayer

Customer Service.

The display
goes blank when
the meter is
turned on
.

1. You may have

accidentally pressed

the

button.

2. Meter has turned

off after 3 minutes

without use or after

10 minutes of no use

while waiting for you

to apply blood or

control solution.

1. Press the

button again.

2. This is a normal expected action

that saves battery life. Press

to turn on again.

Meter will not
turn on
.

1. Battery life has ended.

2. Battery holder is

loose or battery is

installed with the

+

side down.

1. Replace the battery (see page 35).

2. Check the battery and battery

holder to be sure they are

tightly in place and correctly

installed (see page 37). If the

display is still blank, replace the

battery (CR2032).

Meter fails to
start countdown
after blood is
applied.

1. Not enough blood in

test strip.

2. There may be debris

inside the meter.

3. Meter failure.

1. Discard the used test strip. Retest

using a round drop of blood.

2. Inspect the inside of the meter

(see page 34) and remove

debris if present. Retest.

3. Contact Bayer Customer Service.



www.BayerDiabetes.com