BNC 1120 - 1pS Time Interval Counter User Manual

Page 23

Advertising
background image

Model 1120 - 1pS Time Interval Counter

NUT 012

Ed. 6

Date : 18/01/2011

BERKELEY NUCLEONICS

Page

-23 / 36 -

RUN

Syntax:

RUN DELAY|FREQ|COUNT

RUN?

Form:

Set & Query

Description:

Start measure or ask if measuring.

Command

Parameter:

DELAY start a delay measurement

FREQ start a frequency measurement

COUNT start a count measurement

Query response

:RUN YES measurement running

:RUN CAL calibration running

:RUN NO no measurement running

Example:

RUN DELAY

RUN? => :RUN YES

SAMPLES

Syntax:

SAMPLES <samples>

SAMPLES?

Form:

Set & Query

Description:

Set or query the measurement samples.

Command

Parameter:

<samples> number of samples

Specified limit

1 <= samples <= 1000

Example:

SAMPLES 500

SAMPLES? => :SAMPLES 500

Advertising