Model 1120 - 1pS Time Interval Counter

NUT 012

Ed. 6

Date : 18/01/2011

BERKELEY NUCLEONICS

Page

-9 / 36 -

3. O

PERATING INFORMATION

3.1.

Principle:

3.1.1. block diagram

Input

control

Timing

controller

Time counter

Interface Controller

155 MHz

Clock

Calibrator

Linear

interpolator

Cal

Ethernet

Linear

interpolator

Clock

converter

10 MHz

Start

Stop

Keyboard

display

Gate

3.1.2. Time interval mode

In timing interval mode the BNC 1120 measures the time between a Start and Stop
pulse. The timing information is obtained by a processing that combines the two following
information:

Coarse information provided by a wide range Time counter running at 155.52 MHz

(6.43 ns resolution),

And fine information provided by the two linear interpolators started by the event

on the Start and Stop inputs.

A gate input can be used to enable or disable all Start and Stop events as desired.

The BNC 1120 can make positive or negative time measurements. The start pulse can
occur before or after the stop pulse.
The number of sequence measurement can be controlled by the sequence command
single/repetitive. In repetitive sequence the number of measurement is controlled by the
value write in sample command.