BNC 1120 - 1pS Time Interval Counter User Manual

Page 24

Advertising
background image

Model 1120 - 1pS Time Interval Counter

NUT 012

Ed. 6

Date : 18/01/2011

BERKELEY NUCLEONICS

Page

-24 / 36 -

STAT

Syntax:

STAT CLEAR

STAT? POW|TRIG|CLK|TEMP|SCAN

Form:

Set & Query

Description:

Clear or query the status.

Command

Parameter:

CLEAR clear the trig status

SCAN, this command measure all analog values, convert analog

value in digital value, save the digital value in the interface

controller and refresh the value in the front panel

Query

Parameter

POW return the power levels (+6V, -6V, +3.3V, +1.8V, +12V)

TRIG return the trigger status

CLK return the clock status

TEMP return the equipment temperature

Example:

STAT CLEAR

STAT SCAN

STAT? POW => :STAT POW,6.00,-6.00,3.30,1.80,12.00

STAT? TRIG => :STAT TRIG,1,1

STAT? CLK => :STAT CLK,INTERNAL

STAT? TEMP => :STAT TEMP,45.27