Model 1120 - 1pS Time Interval Counter

NUT 012

Ed. 6

Date : 18/01/2011

BERKELEY NUCLEONICS

Page

-29 / 36 -

5. O

PERATION EXAMPLES

5.1.

Example 1

Realize in local mode a repetitive Time interval measurement.

Initial setup

Turn on the BNC 1120 with switch located on the rear panel
After few second the first line of the menu, BNC 1120 is displayed
The instrument must warm-up at least 20 minutes for maximum accuracy
Test signal

The operation example is performed using a test signal with following characteristics (P

ulse

generator):

Duration:

1000 ns,

Rise time

5 ns,

Level:

3 V,

Repetition time :

100 Hz,

This signal must be coupled with 50 divider then connected to start and stop input of BNC 1120
The length of the START cable = 1 m. T
Setting up the instrument

Use the front panel keyboard to reset the Setup menu as follows

Start trigger

External

Start level

1.50 V

Start polarity

+

Stop trigger

External

Stop level

1.50 V

Stop polarity

+

Samples

1000

Sequence

repetitive

Arming mode

on start

Start the measurement

Display the result line Mean delay / jitter, and press the OK keys to start the measurement
The front panel leds busy must be on and tilt each second
Result must be
Delay = 5 200 ns
Jitter = 8 ps rms
For stopping the measurement press the ESC key