Model 1120 - 1pS Time Interval Counter

NUT 012

Ed. 6

Date : 18/01/2011

BERKELEY NUCLEONICS

Page

-35 / 36 -

6.1.3. Command structure

STAT

Syntax:

STAT CLEAR

STAT? POW|TRIG|CLK|TEMP|POPT|SYN

Form:

Set & Query

Description:

Clear or query the status.

Command

Parameter:

CLEAR: clear the trig status

Query

Parameter

SYN return the synthesis word

01 power fault

02 sync error

04 inhibited

POW return the power levels (+6V, -6V, +3.3V, +1.8V, +12V)

TRIG return the trigger status

CLK return the clock status

TEMP return the equipment temperature

POPT return the optical received power

Example:

STAT CLEAR

STAT? SYN => :STAT SYN,#H0001

STAT? POW => :STAT POW,6.00,-6.00,3.30,1.80,12.00

STAT? TRIG => :STAT TRIG,1,1

STAT? CLK => :STAT CLK,INTERNAL

STAT? POPT => :STAT POPT,4.54

STAT? TEMP => :STAT TEMP,45.27