5 troubleshooting, 1 error search, Troubleshooting 5.1 – Eppendorf Pipette Reference 2 User Manual

Page 17: Error search, 5troubleshooting 5.1 error search

Advertising
background image

Troubleshooting

Eppendorf Reference

®

2

English (EN)

17

5

Troubleshooting

5.1

Error search

Symptom/message Cause

Remedy

Liquid is dripping
from the tip and/or
the dispensed
volume is incorrect.

The tip is loose or the pipette tip is poorly
fitted.

Firmly attach the pipette tips. Use
epT.I.P.S. pipette tips.

Liquid with high vapor pressure and/or
different density.

Wet the tip several times and adjust
the pipette for the liquid used.

Pipetting was too quick.

Move the control button slowly.

The tip is withdrawn from the liquid too
quickly.

Withdraw the tip slowly from the
liquid with a time delay (approx. 3
seconds).

Liquid aspirated with blow-out and
dispensed with blow-out.

Repeat dispensing correctly.

Piston dirty or damaged.

Clean piston, grease slightly and/or
replace.

The tip cone is damaged.

Replace the lower part or channel.

The O-rings of the tip cones are
damaged.

Replace the O-rings (only 100

μL,
300

μL

multi-channel).

The control button
jams and does not
move smoothly.

• The piston is dirty.

• The seal is dirty.

• Seal defective.

• The pipette is blocked.

Clean the lower part.

Replace the protection filter on 2

mL

to 10

mL

models.

The adjustment
seal has been
removed; the
adjustment display
has been changed.

The pipette has been adjusted for a
different liquid.

Adjust the pipette for the liquid used.

No spring-loading
action of the tip
cone when pipette
tips are inserted.

Spring-loading action is blocked by a
locking ring.
Spring loading is switched off
(multi-channel).

Remove the locking ring again.

Switch on spring loading.

Use of a 2

mL to 10

mL pipette.

No remedy available. The tip cone of
these sizes does not have spring-loading
action.

Advertising