7.4

Test conditions

Test conditions and test evaluation match in compliance with ISO

8655,

Part

6. Fine balance with

evaporation protection inspected by the office of weights and measures

• Number of determinations:10

• Use of water in accordance with ISO 3696

• Inspection at 20

°C – 27

°C

±0.5

°C

• Dispensing on the tube wall

7.5

Ambient conditions

Technical specifications subject to change.

Temperature range

Relative humidity

Storage without transport
packaging

-5

°C — 45

°C

10

% — 95

%

Operating conditions

5

°C — 40

°C

10

% — 95

%