C.B.S. Scientific SG Series User Manual

Page 17

Advertising
background image

C.B.S. œ Scientific

17

Adjustable Height Sequencer

4.3 Electrophoresis

1. After loading the samples, attach the reservoir safety cover. Connect the DC power leads to

the unit then to the power supply with the proper polarity. Make sure the black leads are
connected to the black cathodes (-) and the red leads are connected to the red anode (+).


2. Turn on the power supply and set the voltage or wattage to the proper setting for the gel. Be

sure to follow manufacturer's instructions. (See Table 5.1 for recommended DC power
settings). As a general rule, the applied voltage is approximately 50-60V per cm of gel.
Monitor temperature using the Temperature strip (Cat. # EGT-100).


3. Monitor the progress of electrophoresis by following migration of a marker dye front or by any

other preferred method.


4. When electrophoresis is complete, TURN POWER SUPPLY OFF. Disconnect both DC

power cords, first from the power supply.

Advertising
This manual is related to the following products: