Adjustable Height Sequencer

Section 6
Maintenance of Equipment

6.1

Care and Handling

The components of the Sequencer are fabricated from acrylic, Delrin®, Polycarbonate, glass and
anodized aluminum. As with any laboratory instrument, adequate care ensures consistent and
reliable performance.

After each use, rinse buffer reservoirs, combs, spacers, clamps and glass plates with deionized
water. Wipe dry with a soft cloth or paper towel, or allow to air dry. Whenever necessary, all
components may be washed gently with water and a non abrasive detergent, and rinsed and
dried as above. Never use abrasive cleaners, window sprays or scouring pads to clean the
components, as these cause damage to the plastic surfaces.


Additional

cautions:

• Do not autoclave or dry-heat sterilize the apparatus or components.

• Do not expose the apparatus or components to phenol, acetone, benzene, halogenated

hydrocarbon solvents or alcohols.

• Avoid prolonged exposure of the apparatus or components to UV light.

6.2 Maintenance

The following simple, routine inspection and maintenance procedures will help ensure both the
safety and the performance of the Nucleic Acid Sequencers. For ordering information of
replacement parts call 1-858-755-4959.

• Inspect electrical connections and power cords regularly. If the banana plug connectors

to either the unit or the power supply do not exhibit reasonable friction or can be rocked
easily, replace the plugs or power cords.

• If power cords show any signs of wear or damage (e.g. cracks, nicks, abrasions, or

melted insulation), replace them immediately.

Examine the banana jacks to ensure that they are tight and free of corrosion. If banana
jacks are corroded, they should be replaced.