CiDRA SONARtrac HD GVF-100 User Manual

Page 101

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 11-4

20675-02 Rev 01

temperature sensor will be used to input pressure to the transmitter
press ‘ENTER’ and use the

key to select either ‘Sensor #1’ or

Sensor #2’. (‘Sensor #1’ or ‘Sensor #2’ refers to the sensor input
the temperature transducer is wired to on the terminal block within
the transmitter box.) Once the selection is made, press ‘ENTER’.

Pressing the

key will next display ‘→Altitude’ on Line 2 and the

selected value above (or below) sea level on Line 4. To change
the Line 4 value press ‘ENTER’ and use the arrow keys to enter
the new value. Note: if an external pressure transducer that reads
pressure in ‘absolute’ values (e.g. PSIa, BARa, or KPaa is used an
altitude correction is not necessary.

Pressing the

key will next display ‘→Calibration’ on Line 2.

Press the ‘ENTER’ key and ‘

→C0’ will appear on Line 3 and a

numerical value on Line 4. Press the ‘ENTER’ key and use the
arrow keys

to enter the calibration factors that will accompany the

sensor. Once the ‘C0’ value is entered press the ‘ENTER’ key to
store that value to memory. Line 3 will show ‘

→C0’; press the

key to move to ‘

→C1’, press ‘ENTER’, and use the arrow keys to

enter ‘C1’ values. Once the values are entered, press ‘ENTER
and press the

key to move to ‘→C2’ and enter those values and

press ‘ENTER’ and the ‘BACK’ key.

The display will show ‘

→Calibration’ on Line 2. Press the key to

scroll to

→Flow Direction’ on Line 2. If it is necessary to change

the flow direction (if the sensor was installed with the Flow
Direction arrow on the band opposite the actual flow direction
within the pipe or the process flow has changed direction) press
the ‘ENTER’ key and toggle the

↑ or ↓ to change flow direction.

Once changed press the ‘ENTER’ key.

The display will show ‘

→Flow Direction’ on Line 2. Press the

key to scroll to ‘

→Op Mode’ on Line 2. Press the ‘ENTER’ key

and the current operational mode will be displayed on Line 4 of the
display. Press the

key to and scroll to ‘Flow/GVF/SOS’ and then

press the ‘ENTER’ key to set that selection.

Press the

key to scroll to ‘→Set Date/Time’ on Line 2. Press the

ENTER’ key and the current saved date and time will be displayed
on Line 4 of the display. Use the arrow keys (

and to scroll

through the values and

and to change character) to enter the

date and time. If possible, synchronize the date and time entered
to that of the process control data system. Note: Time is in 24-
hour format. Once the date and time have been set press the
ENTER’ key followed by the ‘BACK’ key.

At this point ‘Basic Config’ is displayed on Line 1 of the display.
The user can re-enter this menu if desired by pressing the
ENTER’ key or by pressing the

or arrows to move to other

Level 1 menus (see Table 12).