SENSOR REMOVAL PROCEDURE

If removal of the sensor is required, the following procedure should be
followed.

1. Specific details regarding the removal of the SONARtrac sensor

should be discussed with a CiDRA representative prior to removal.

2. Obtain Hot Work Permit if required.

3. Turn off electrical power to the system. Follow lock-out / tag-out

procedures as required.

4. Open the transmitter door using a screwdriver and remove the

sensor band shorting plug (stored in the transmitter) if there is one.

5. Unlatch and disconnect the sensor to transmitter cable connector

from the sensor cover. Close cable connector cover on sensor
head cover.

6. Protect cable connector from dirt, water, etc. (example, wrap it in

a protective plastic bag).

7. Remove the rain seal(s) or boot gasket fenders (stainless steel

covers) from the sensor head (if equipped).

8. Open the sensor cable access panel located on the cover by

loosening the six screws (they should not be fully removed from
the cover as they are captive screws).

9. Under the access panel removed above, disconnect the sensor

band to cover electrical connector by unlatching the connector
locks and sliding the connector back as shown below.