Copyright © 2006 CiDRA Corporation

Page 7-6

20675-02 Rev 01

Position the SONARtrac

TM

sensor band assembly on the pipe with the

polyimide film (amber colored) against the the compliant layer.
Important:

The attachment rails on an HD sensor band must be

installed on the top of a horizontal flowing pipe. Slide the alignment
pins on the attachment rail through their mating holes on the opposite
attachment rail.

If possible, orient the flow direction arrow on the sensor assembly with
the direction of flow within the pipe. Note: If this is not possible due to
installation constraints, e.g. access to sensor fasteners, install
opposite to flow direction. The transmitter must be re-configured to
“reverse flow” during its set up as detailed in the Transmitter Startup &
Operation Section of the manual.

Figure 7

Sensor Band Screw and Alignment Pins

Carefully start threading the screws into their screw holes (avoid cross
threading) by using the hex driver until each screw is engaged about 2
turns. The following table provides guidance for selecting the proper
hex tool and spacer gauge.

Table 4

Gauge Block and Screw Size Table

Sensor Band P/N

Spacer Gauge

P/N

Socket Head Screw

Hex Size (inch)

Band Attachment Rail

Size (Ref)

20380- ALL SIZES

20143-01

7/64

1/8 x 3/8

20409- ALL SIZES

20143-02

7/64

1/8 x 3/8

20690- ALL SIZES

20143-04

5/32

1/4 x 1/2

20686- ALL SIZES

20143-04

5/32

1/4 x 1/2

Compliant sheet

Sensor attachment screw
assembly (typical 9 places)

Alignment pins
(typical 2 places)

Flow direction arrow

Attachment rails

Sensor band shorting plug

Advertising