Copyright © 2006 CiDRA Corporation

Page 10-8

20675-02 Rev 01

Flow Rate’, ‘Flow Rate %’, ‘SOS’, and ‘GVF’ generate a different
pulse output than ‘Total Flow’ (‘Totalizer’). Flow Rate, Flow Rate %,
SOS, and GVF will output a pulse frequency based on the current
measurement. The pulse width will vary in order to maintain a 50%
duty cycle pulse train.

Figure 64

Speed Of Sound and Gas Volume Fraction Pulse Output

Total Flow’ (or Totalizer) will output a number of pulses based upon
the total gallons (or flow units) counted during the previous update
interval (i.e. the display update time is defaulted to 2 seconds). In
effect, the Total Flow measurement is a pulse counter, updated at the
same time as the display.

Figure 65

Total Flow Pulse Output

SOS

Time (sec)

500 fps / 100 = 5 Hz

1000 fps / 100 = 10 Hz

300 fps / 100 = 3 Hz

Multiplier = 100

0

2

4

6

Pulse Width Varies

Total Flow (Totalizer)

Time (Update Rate = 2 sec)

0

2

4

6

(1500 gpm /10) / (60s/2s) = 5

(3000 gpm /10) / (60s/2s) =10

(900 gpm /10) / (60s/2s) = 3

Multiplier = 10

Pulse Width