A2 vf & gvf-100 performance specifications – CiDRA SONARtrac HD GVF-100 User Manual

Page 118

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page A-6

20675-02 Rev 01

A2

VF & GVF-100 PERFORMANCE SPECIFICATIONS

A2.1 Turn-on

Time

30 minutes to rated accuracy from power up

25 seconds from power interruption

A2.2 Start-up

Time

25 seconds from zero flow

A2.3

Low / High Flow Cutoff

Adjustable between 3 and 30 ft/sec. At process flow rates below and
above these values, the output will register a ‘<min flow’ or’>max flow’
indication.

A2.4 Flow

Rate

Range

SONARtrac

TM

is capable of measuring signals from process fluids

traveling at velocities between 3 ft/s and 30 ft/s (1.0 m/s to 10 m/s).
This capability is for both forward or reverse flow directions.

A2.5 Flow

Rate

Accuracy

Flow rate accuracy is +/-1.0% of flow rate from 3 to 30 ft/sec (1 to
10m/s).

A2.6

Gas Volume Fraction Range

SONARtrac

TM

is capable of determining the Gas Volume Fraction in

fluids between 0% and 20% entrained air.

A2.7

Gas Volume Fraction Accuracy

Accuracy of +/- 5% of reading within the 0.01% to 20% entrained
gas/air range when used with an in-line process pressure reading.

A2.8 Repeatability

+/- 1% of reading

A2.9 Update

Rate

2 seconds