7.3

Sensor Installation Guidelines

The following are general installation guidelines and recommendations
for installing a SONARtrac

TM

sensor.

Where necessary, get a Hot Work Permit prior to installation of
system.

Select locations with well-developed flow profiles.

Avoid installation locations directly after piping configurations that
cause flow jetting.

Install flow sensor upstream of control valves, “T”s, orifice plates,
and any other severe source of flow disturbance.

Locate the sensor upstream of pipe taps such as those used for
temperature and pressure sensors.

Good piping practices are required near flanges. This includes
good alignment of pipes, and properly sized and installed gaskets
that do not disturb the flow profile.

Contact your local distributor or CiDRA Technical Support if you
have questions.

Table 3 lists the recommended installation distances from flow
disturbances. These recommendations apply to flow measurement
installations.

Feature

Upstream

pipe

diameters

Downstream

pipe

diameters

90 Degree Elbow

15

5

Double Elbow in plane

25

5

Double Elbow out of plane

40

5

Diffuser (expansion)

30

5

Reducer 15

5

Control Valve

40

10

Flange 2

2

Table 3

Recommended Distances From Flow Disturbances

7.4

Process Piping Considerations

The SONARtrac

TM

sensor head assembly mounts on the process

pipe. There is no need for breaking any process connections or for
shutting down the process. The sensor head must be installed in a
location that ensures a full pipe during operation.