7.5.1

Sensor Band Short Test

Shorting of the sensor band to the process pipe may cause signal
interference or electrical faults in the system in some instances. The
sensor band must be electrically isolated from the process pipe.

Use an ohm meter and verify the sensor band is isolated from the
pipe. Measure the resistance between the sensor rails and the pipe to
ensure there is no continuity between the sensor band and the
process pipe. If the band is shorted identify where the short is located
and eliminate the short. For example, if a sensor screw is shorting to
a pipe weld bead, reposition the sensor band, or lightly file the weld
bead to eliminate the interference.

7.5.2

Sensor Band Thermal Barrier Installation

Install the sensor band thermal barrier if one was included with the HD
VF-100 System. Refer to the figure below.

1. Align the slit on the thermal barrier with the sensor band to pre-

amplifier cable.

2. Wrap the thermal barrier over the sensor attachment rails.

3. Continue to wrap the thermal barrier around the sensor band.

4. Seal at the Velcro strips and install the straps through the D-rings

on the thermal barrier.

5. Retain the sensor ban to pre-amplifier cable in the Velcro retention

loop.

Figure 13

Thermal Barrier Installation Over Sensor Band

Mounting rails on top of
pipe covered with
thermal barrier

Velcro strip for sealing
thermal barrier

Sensor band to pre-
amplifier cable in slot on
thermal barrier

Sensor band to pre-
amplifier cable in
retention loop

Thermal barrier
straps (both sides)