1 transmitter output connections (section #1), Comment, Figure 36 transmitter output terminals -5 – CiDRA SONARtrac HD GVF-100 User Manual

Page 46: Table 5 list of transmitter signal outputs -5, Warning

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 8-5

20675-02 Rev 01

8.4.2.1 Transmitter

Output Connections (Section #1)

The following figure provides a close-up of the transmitter output
terminals (Section #1 of the terminal board) with their functions listed
in the following table. These outputs can be connected as appropriate
to permit communications between the transmitter and other
equipment.

Figure 36

Transmitter Output Terminals

Terminal Label

Type

Comment

CUR 1

4-20mA #1 Output

Internal (self) or external (loop) powered,
HART communication capable

CUR 2

4-20mA #2 Output

Internal (self) or external (loop) powered

PULSE

Pulse output

Solid-state relay closure

ALARM

Alarm output – High /
Low Output

Solid-state relay closure

COMM

RS-485 or 232 Output

User configured

SHD ---

Shield

Connections

Table 5

List of Transmitter Signal Outputs

User supplied data output cable (size range AWG 30 to AWG 10) is
installed through a cable gland in the furthest left hole on the
transmitter housing and attached to the appropriate terminal block
connection points. The cable gland is sized for a 3/4 inch NPT fitting
(1-1/16 inch hole).

WARNING

For Class I, Division 2 applications, Power Entry and
Inputs/Outputs must be installed in accordance with Article
501.10(B)(1) of the National Electrical Code ANSI/NFPA 70:2005