SEnSoR CoMPARISon/CRoSS SEnSITIvITY

Typical Cross Sensitivity of H

2

S Sensor (0-20 ppm / 0-50 ppm / 0-100 ppm)

Gas

Concentration

Reading

Carbon monoxide

300 ppm

≤ 2 ppm

Sulfur dioxide

5 ppm

~ 1 ppm

Nitric oxide

35 ppm

< 0.7 ppm

Hydrogen

10000 ppm

≤ 10 ppm

Nitrogen dioxide

5 ppm

~ –1 ppm

Factory Mutual Performance Approved Electrochemical Gas Sensors

* Time to reach percentage of final reading when gas concentration equal to full scale is applied to sensor.
** Background concentrations of ammonia may shorten lifetime of sensor.
*** Sensor approved for oxygen depletion (< 21% V/V) only.

Gas

Range

Response Time*

Accuracy of

Reading

(Whichever

is Greater)

Operating

Temperature Range

Zero drift

Performance

Approved Standard

Hydrogen Sulfide

(H

2

S)

0-20 PPM

T50 = 10 Sec.
T90 = 23 Sec.

±2 ppm or

±10% of Reading

–40°C to +50°C

± 1 ppm/Mo.

ISA 92.00.01

Hydrogen Sulfide

(H

2

S)

0-50 PPM

T50 = 10 Sec.
T90 = 23 Sec.

±2 ppm or

±10% of Reading

–40°C to +50°C

± 1 ppm/Mo.

ISA 92.00.01

Hydrogen Sulfide

(H

2

S)

0-100 PPM

T50 = 12 Sec.
T90 = 28 Sec.

±2 ppm or

±10% of Reading

–40°C to +50°C

± 2 ppm/Mo.

ISA 92.00.01

Ammonia (NH

3

)

0-100 PPM**

T50 = 24 Sec.
T90 = 65 Sec.

±4 ppm or

±10% of Reading

–20°C to +40°C

± 2 ppm/Mo.

FM6340

Ammonia (NH

3

)

0-500 PPM**

T50 = 30 Sec.

T90 = 120 Sec.

±4 ppm or

±10% of Reading

–20°C to +40°C

± 10 ppm/Mo.

Det-Tronics Verified

(CSA Exd)

Oxygen (O

2

)

0-25% V/V***

T20 = 7 Sec.

T90 = 30 Sec.

< 0.5% V/V

–20°C to +50°C

< 2 %/Mo.

BS EN 50104

Carbon

Monoxide (CO)

0-100 PPM

T50 = 15 Sec.
T90 = 40 Sec.

±5 ppm or

±10% of Reading

–20°C to +50°C

± 2 ppm/Mo.

ISA 92.00.01

Carbon

Monoxide (CO)

0-500 PPM

T50 = 12 Sec.
T90 = 25 Sec.

±5 ppm or

±10% of Reading

–20°C to +50°C

± 9 ppm/Mo.

ISA 92.00.01

Sulfur Dioxide

(SO

2

)

0-20 PPM

T50 = 12 Sec.
T90 = 30 Sec.

±0.6 ppm or

±10% of Reading

–20°C to +50°C

± 0.4 ppm/Mo.

Det-Tronics Verified

(CSA Exd)

Sulfur Dioxide

(SO

2

)

0-100 PPM

T50 = 15 Sec.
T90 = 35 Sec.

±0.6 ppm or

±10% of Reading

–20°C to +50°C

± 0.4 ppm/Mo.

Det-Tronics Verified

(CSA Exd)

Chlorine

(Cl

2

)

0-10 PPM

T50 = ≤14 Sec.
T90 = ≤34 Sec.

±0.6 ppm or

±10% of Reading

–20°C to +50°C

< 0.2 ppm/Mo.

FM6340

Hydrogen

(H

2

)

0-1,000 PPM

T50 = 8 Sec.

T90 = 60 Sec.

±50 ppm or

±10% of Reading

–20°C to +40°C

± 20 ppm/Mo.

Det-Tronics Verified

(CSA Exd)

Nitrogen Dioxide

(NO

2

)

0-20 PPM

T50 = 7 Sec.

T90 = 31 Sec.

±2 ppm or

±10% of Reading

–20°C to +40°C

± 0.1 ppm/Mo.

Det-Tronics Verified

(CSA Exd)