9.1

95-8616

Typical Cross Sensitivity of NH

3

Sensor (0-100 ppm)

Gas

Concentration

Reading

Alcohols

1000 ppm

0 ppm

Carbon dioxide

5000 ppm

0 ppm

Carbon monoxide

100 ppm

0 ppm

Hydrocarbons

% Range

0 ppm

Hydrogen

10000 ppm

0 ppm

Hydrogen sulfide

20 ppm

~ 2 ppm

1

1

Short time gas exposure in minute range.

Typical Cross Sensitivity of NH

3

Sensor (0-500 ppm)

Gas

Concentration

Reading

Alcohols

1000 ppm

0 ppm

Carbon monoxide

100 ppm

0 ppm

Chlorine

5 ppm

0 ppm

Nitrogen dioxide

10 ppm

0 ppm

Sulfur dioxide

20 ppm

–40 ppm

Hydrogen

3000 ppm

0 ppm

Hydrogen sulfide

20 ppm

2 ppm

Typical Cross Sensitivity of CO Sensor (0-100 ppm / 0-500 ppm)

Gas

Concentration

Reading

Hydrogen sulfide

15 ppm

~ 45 ppm

Sulfur dioxide

5 ppm

~ 2.5 ppm

Nitric oxide

35 ppm

~ 10 ppm

Chlorine

1 ppm

–1 ppm

Hydrogen

100 ppm

< 40 ppm

Nitrogen dioxide

5 ppm

~ –3 ppm

Typical Cross Sensitivity of NO

2

Sensor (0-20 ppm)

Gas

Concentration

Reading

Alcohols

1000 ppm

0 ppm

Carbon dioxide

5000 ppm

0 ppm

Chlorine

1 ppm

≤ 1 ppm

Nitric oxide

100 ppm

0.4 ppm

Sulfur dioxide

20 ppm

5 ppm

Hydrogen

3000 ppm

0 ppm

Typical Cross Sensitivity of SO

2

Sensor (0-20 ppm / 0-100 ppm)

Gas

Concentration

Reading

Carbon monoxide

300 ppm

< 1 ppm

Nitric oxide

50 ppm

0-5 ppm

Nitrogen dioxide

6 ppm

< –10 ppm

Hydrogen Sulfide

25 ppm

< 0.1 ppm

Chlorine

5 ppm

< –2 ppm

Ammonia

20 ppm

0 ppm

Hydrogen

400 ppm

< 1 ppm

Hydrogen Cyanide

10 ppm

< 5 ppm

Acetylene

10 ppm

< 30 ppm

Ethene

50 ppm

< 45 ppm

Typical Cross Sensitivity of Cl

2

Sensor (0-10 ppm)

Gas

Concentration

Reading

Carbon monoxide

300 ppm

0 ppm

Hydrogen sulfide

15 ppm

~ –7.5 ppm

Sulfur dioxide

5 ppm

0 ppm

Nitric oxide

35 ppm

0 ppm

For details on other interfering gases, please contact Detector Electronics Corp.