Startup procedure, Troubleshooting, Maintenance – Det-Tronics PIRVOLHV Volumetric Infrared Hydrocarbon Gas Process Monitor for Heavy Hydrocarbons User Manual

Page 10

Advertising
background image

8

3 .1

95-8632

STARTUP PROCEDURE

NOTE

Caution should be exercised when applying this
and any electronic device to explosive vapor
extraction and measurement applications.
Typical hazards in these applications include
small fitting vapor leaks, resulting in dangerous
gas concentrations near or within the system
enclosure. Any spark or arc may result in a fire
or explosion in these conditions. Always inspect
all fittings for leakage before commissioning, and
always remove power from all electronic devices
before performing maintenance on the vapor
extraction system tubing, fittings, and hardware.

1. Inhibit the output loads that are actuated by the

system to prevent activation of these devices.

2. Check that the PIRVOLHV has been wired

properly.

3. Apply power to the system and allow the

PIRVOLHV to operate for a minimum of 2 hours,

then check zero and verify gas response.

4. Place the system in normal operation by

reactivating the output loads.

TROUBLESHOOTING

Refer to Table 2 to isolate and correct malfunctions with

the PIRVOLHV.

MAINTENANCE

It is recommended to have spare IR Modules on hand

(see “Spare Parts” section). Use Table 2 to isolate and

correct malfunctions.

IMPORTANT MAINTENANCE NOTES

• Hydrocarbon-based grease emits hydrocarbon
vapors, which will be measured by PIRVOLHV,
resulting in elevated gas concentration
readings.

Use only silicone-based grease (not

hydrocarbon-based grease) when lubricating
threads on the PIRVOLHV and associated
termination box.
A suitable grease is listed under
“Spare Parts” in the Ordering Information section
of this manual.

• It is recommended to keep spare IR modules
(See “Spare Parts” section) for field replacement
in the event of a malfunction.

Table 2—Troubleshooting Table

Current Level

Status

Corrective action

2.4 to 3.9 mA

Negative Zero Indication

This fault can be caused by the presence of

background gas during calibration. Ensure that

background gas is not present and recalibrate the

unit. If fault does not clear, perform disassembly

and cleaning procedure, then recalibrate. If fault

still does not clear, replace electronics assembly.

1.6 mA

Calibration fault

Repeat zero calibration. If the condition persists,

replace electronics assembly.

1.0 mA

Fouled optics

Perform disassembly and cleaning procedure,

then recalibrate.

0.8 mA

+24 Vdc line low (less than +17.5 Vdc)

Ensure that input voltage is correct and that power

connections are good. If fault does not clear,

replace the electronics assembly.

0.6 mA

Calibrate input active at power-up

Ensure that calibration line is not shorted and that

the calibration switch is open. If fault does not

clear, replace the unit.

0.4 mA

Active channel fault

Replace electronics assembly.

0.2 mA

Reference channel fault

Replace electronics assembly.

0.0 mA

CPU system fault, warmup

Ensure that power is applied and that the warmup

period is complete (1 minute). If fault does not

clear, replace the unit.

Advertising