Det-Tronics PIRVOLHV Volumetric Infrared Hydrocarbon Gas Process Monitor for Heavy Hydrocarbons User Manual

Page 11

Advertising
background image

9

3 .1

95-8632

DISaSSEMBLy aND CLEaNINg pROCEDURE
The PIRVOLHV should be inspected periodically to ensure

that its performance is not impaired by fouled optics or by

clogging of the sample draw cup or hydrophobic filter.

Inspection and/or periodic maintenance involves three

different areas.

IMPORTANT

Remove power before disconnecting and
removing the PIRVOLHV for maintenance.

Sample Draw Cup.

Perform a visual inspection of the

sample draw cup and associated plumbing, and clean

as required. External filtration of sample stream is

recommended.

hydrophobic Filter.

The flow of gas through the

filter can be inhibited by an accumulation of mud or

other contaminants. To inspect the hydrophobic filter,

disassemble the PIRVOLHV as described below. If the

filter appears to be fouled, replace it.

Optics.

Cleaning of the optical surfaces is required only

if an optical fault is indicated (1.0 mA current output signal

from the PIRVOLHV). This procedure is most easily

accomplished on a bench.

Required materials: Clean, flat work surface, foam tipped

swabs (no cotton), isopropyl alcohol, screwdriver.

CAUTION

The PIRVOLHV contains semiconductor devices
that are susceptible to damage by electrostatic
discharge. An electrostatic charge can build
up on the skin and discharge when an object is
touched. Therefore, use caution when handling
the device, taking care not to touch electronic
components or terminals. If the electronics
assembly is removed, it should be placed in an
anti-static bag or box while stored or transported.
A static safeguarded work area is highly
recommended (if available) for disassembly and
cleaning of the PIRVOLHV.

1. Remove the sample draw cup from the front of the

PIRVOLHV by pulling it straight off the electronics

assembly (no tools needed). See Figure 12.

2. Perform a visual inspection of the sample draw

cup and associated plumbing, and clean as

required. Also inspect the two O-rings inside the

sample draw cup and clean, lubricate or replace

as required.

3. Loosen the two captive screws on top of the

mirror assembly (Figure 12) and remove the

mirror assembly and hydrophobic filter from the

electronics assembly.

4. Thoroughly douse the interior of the mirror

assembly as well as the foam tipped swab

with isopropyl alcohol. Use the swab to gently

cleanse the surfaces of the reflecting mirrors

inside the mirror assembly. After cleaning with

the swab, flush out the mirror assembly using

a liberal amount of isopropyl alcohol. Tip the

mirror assembly with mirror openings downward

to remove accumulated isopropyl alcohol and

particle contaminants. Repeat the alcohol flush

to remove any remaining contaminants. Allow the

mirror assembly to air dry in a dust-free location.

IMPORTANT

Do not insert any sharp object into the mirror
assembly. Scratching of the mirrors will void the
PIRVOLHV warranty. Do not use cotton tipped
swabs or buds as they are likely to leave fiber
residue.

5. Clean the two windows on the electronics

assembly using the procedure described above.

When the mirror assembly and electronics

assembly are thoroughly dry, proceed with

re-assembly.

6. Slide a new hydrophobic filter over the end of the

mirror assembly, then carefully seat the mirror

assembly and hydrophobic filter against the base

of the electronics assembly. Be careful not to

crumple or fold the hydrophobic filter.

7. Tighten the two captive screws on the top of the

mirror assembly. Tighten the screws evenly. Do

not over-tighten (6 inch-pounds recommended).

8. Slide the sample draw cup back onto the

PIRVOLHV.

9. Check for leaks around the PIRVOLHV sample

draw cup.

Advertising